phone Phone
+622129330305
menu

CV. Alkestron Bagusjaya Mandiri

sql
GENESYS 150 UV-Visible Spectrophotometer
Hover to Zoom

Jual GENESYS 150 UV-Visible Spectrophotometer



Spesifikasi GENESYS 150 UV-Visible Spectrophotometer

Specifications:

 

Baseline Flatness± 0.002 A
CompartmentsAccommodates optional accessories
8-position cell changer
4-position cell changer (long path cells)
Peltier thermostatted cell holder (20 °C - 60 °C)
Sipper accessory
Fiber optic probe coupler
ConnectionsSingle USB-A supports flash memory devices for method and data storage
Duplex USB-A on side supports connection to a Windows™ computer running optional remote control software, keyboard, and mouse;
Export data to network or PC via USB, Ethernet, or Wi-Fi
DescriptionGENESYS™ 150 UV/Visible Spectrophotometer
Detector TypeDual Silicon Photodiodes
Dimensions, 
(L x W x H)
35.5 x 38.5 x 19.5 cm
Display7-inch color touchscreen, tiltable, high definition, 800 x 1280 pixels
Drift< 0.0010 A/Hr (At 500 nm after 1 hour warmup)
KeypadTouchscreen
LampXenon Flash Lamp (> 5 years typical, 3 years guaranteed)
Noise(RMS at 500 nm 60 consecutive measurements)
</= 0.00020 A @ 0 A @ 260 and 500 nm
</= 0.00030 A @ 1 A @ 260 and 500 nm
</= 0.00040 A @ 2 A @ 260 and 500 nm
Optical DesignDual beam
PhotometricInstrument
Accuracy
± 0.002 A @ 0.5 A
± 0.004 A @ 1.0 A
± 0.008 A @ 2.0 A
Display- 3 A to + 5 A
Range- 2 A to + 3.5 A
Repeatability± 0.001 A @ 1 A (measured at 1.0 A at 546 nm)
Power RequirementsExternal AC to DC converter, Voltage and Frequency (Hz) are selected automatically, 100 to 240 V, 50 to 60 Hz
PrinterSnap-on printer (optional—purchase separately)
Resolution0.2 nm. 0.5 nm, 1 nm, 2 nm, 5 nm
Spectral Bandwidth5 nm
Stray Light< 1.0% T @ 198 nm (KCL)
< 0.05 % T @ 220 nm (Nal)
< 0.03 %T @ 340 nm (NaNO2)
WavelengthAccuracy± 0.5 nm
Range190 nm to 1100 nm
Repeatability< ± 0.2 nm
Scan SpeedSlow, medium, and fast — up to 1600 nm/min
Weight7.5 kg




Alat Laboratorium Umum

Untuk keterangan lebih lanjut, silahkan download PDF ini :